Keithley 4200-SCS cannot be found via the visadevlist command.

5 vues (au cours des 30 derniers jours)
Nathaniel Hernandez
Nathaniel Hernandez le 11 Sep 2022
Réponse apportée : Andrei le 22 Sep 2022
Hi,
I have a Keithley 4200-SCS connected to my machine via an NI USB/GPIB interface. I also have the newest version of Matlab (2022a).
The visadevlist command does not populate with my device.
I have the NI Max software installed and can communicate via the NI VISA > VISA Interactive Control, by manually inputting commands. I was also able to communicate just fine with the older gpib functions over Matlab, but since they are being deprecated, I was hoping to switch over to the newer visadev commands.
I think the issue is that the Keithley 4200-SCS has it's own older command set instead of the 'newer' standard SCPI command set. When I run the visadevlist command, I see on the keithley that it got an '*IDN?' command, but it errors out since that's not a command it knows; It instead uses the 'ID' command.
Thanks for all the help.

Réponses (1)

Andrei
Andrei le 22 Sep 2022
Hi Nathaniel -- please refer to this bug report for information on which release this issue is resolved in:

Catégories

En savoir plus sur Instrument Control Toolbox Supported Hardware dans Help Center et File Exchange

Produits


Version

R2022a

Community Treasure Hunt

Find the treasures in MATLAB Central and discover how the community can help you!

Start Hunting!

Translated by